Determination of 17beta-oestradiol and testosterone in bovine serum with gas chromatography-mass spectrometry

Warianty tytułu

Języki publikacji

EN

Abstrakty

EN
A simple and sensitive method was developed for determination of 17β-oestradiol and testosterone in bovine serum. The hormones were analysed and quantified as heptafluorobutyric anhydride derivatives by gas chromatography-mass spectrometry after extraction from serum without any cleanup step. The method was validated according to the requirements listed in the Commission Decision 2002/657/EC. Decision limit (CCα) and detection capability (CCβ) were based on the detection of the most abundant ion and the values: 0.011 µg L⁻¹ and 0.018 µg L⁻¹ for 17β-oestradiol were obtained, respectively . The repeatability for the validation level 0.040 µg L⁻¹ was 19.2% and within laboratory reproducibility - 25.8%. For testosterone, CCα and CCβ were 0.056 µg L⁻¹ and 0.096 µg L⁻¹ alternately with repeatability and reproducibility below 12%. The values of the detection limits of 17β-oestradiol and testosterone were found to be below the recommended concentration, set at 0.1 µg L⁻¹ and 0.5 µg L⁻¹, respectively. The method was successfully applied in the official control of hormone residues in Poland.

Słowa kluczowe

Wydawca

-

Rocznik

Tom

55

Numer

4

Opis fizyczny

p.755-759,fig.,ref.

Twórcy

autor
  • Department of Pharmacology and Toxicology, National Veterinary Research Institute, 24-100 Pulawy, Poland
autor

Bibliografia

  • 1. Aksglaede L., Juul A., Leffers H., Skakkebæk N.E., Andersson A.M.: The sensitivity of the child to sex steroids: possible impact of exogenous estrogens. Hum Reprod Update 2006, 12, 341-349.
  • 2. Buisson C., Hebestreit M., Weigert A.P., Heinrich K., Fryc H., Flenker U., Banneke S., Prevost S., Andre F., Schaenzer W., Houghton E, Le Bizec B.: Application of stable carbon isotope analysis to the detection of 17 β-estradiol administration to cattle. J Chromatogr A 2005, 1093, 69-80.
  • 3. Casademont G., Perez B., Garcia Regueiro J.A.: Simultanos determination, in calf urine, of twelve anabolic agents as heptafluorobutyryl derivatives by capillary gas chromatography-mass spectrometry. J Chromatogr В 1966, 686, 189-198.
  • 4. Commission Decision 2002/657/EC, Off J Eur Commun 2002, L 221, 8.
  • 5. Council Directive 2003/74/EC, Off J Eur Commun 2003, L 262, 17.
  • 6. Council Directive 96/22/EC, Off J Eur Commun 1996, L 125, 3.
  • 7. Courant F., Antignac J.P., Maume D., Monteau F., Andersson A.M., Skakkebæk N., André F., Le Bizec B.: Exposure assessment of prepubertal children to steroid endocrine disrupters. Analytical strategy for estrogens measurements in plasma at ultra-trace level. Anal Chim Acta 2007, 586, 105-114.
  • 8. De Brabander H.F.: Pitfalls, methods and criteria in the analysis of residues and contaminants in biological material. Uniwersiteit Gent, Faculteit Diergeneeskunde Veterinaire Volksgezondheid & Voedselveiligheid, Laboratorium voor Chemische Analyse, Salisburylaan, 133 B-9820 Merelbeke. ISBN 90-5864-043-4.
  • 9. Dixon S.N., Russell K.L., Heitzman R.J., Mallinson B.B.: Radioimmunoassay of the aanabolic agent zeranol. V. Residues of zeranol in the edible tissues, urine, faeces and bile of steers treated with Ralgro. J Vet Pharmacol Ther 1986, 9, 353-358.
  • 10. Doyle E., Human Safety of Hormone Implants Used to Promote Growth in Cattle. FRI Briefings, Food Research Institute, University of Wisconsin, Madison, 2000.
  • 11. Draisci R., Palleschi L., Ferretti E., Lucentini L., Cammarata P.J.: Quantitation of anabolic hormones and their metabolites in bovine serum and urine by liquid chromatography-tandem mass spectrometry. J Chromatogr A 2000, 870, 511-522.
  • 12. Eurachem CITAC Guide Quantifying Uncertainty in Analytical Measurement, 2nd ed., 2000.
  • 13. Fedeniuk R.W., Boison J.O., MacNeil J.D.: Validation of gas chromatography-mass spectrometry method for the determination of pg/ml levels of 17β-estradiol and 17β-trenbolone in bovine serum. J Chromatogr В 2004, 802, 307-315.
  • 14. Impens S., De Wasch K., Cornelis M., De Brabander H.F.: Analysis on residues of estrogens, gestagens and androgens in kidney, fat and meat with gas chromatography-tandem mass spectrometry. J Chromatogr A 2002, 955, 281-293.
  • 15. Kootstra P.R., Zoontjes P.W., Van Tricht E.F., Sterk S.S.: Multi-residue screening of a minimum package of anabolic steroids in package urine with GC-MS. Anal Chim Acta 2007, 586, 82-92.
  • 16. Le Bizec B., Antignac J.F., Andre F.: Multidimensional statistical analysis applied to electronic impact mass spectra to determine steroid stereoisomery. Proceedings of Euroresidue V Conference, Noordwijkerhout, The Netherlands, 2004, pp. 325-330.
  • 17. Noppe H., Le Bizec B., Verheyden K., De Brabander H.F.: Novel analytical methods for the determination of steroid hormones in edible matrices. Anal Chim Acta 2008, 611, 1-16.
  • 18. Rambaud L., Monteau F., Deceuninck Y., Bichon E., André F., Le Bizec B.: Development and validation of a multi-residue method for the detection of a wide range of anabolic steroids in hair using gas chromatography-tandem mass spectrometry. Anal Chim Acta 2007, 586, 93-104.
  • 19. Stolker A.A.M., Linders S.H.M.A., Van Ginkel L.A., Brinkman U.A.T: Application of the revised EU criteria for the confirmation of anabolic steroid in meat using GC-MS. Anal Bioanal Chem 2004, 378, 1313-1321.
  • 20. Turpeinen U., Linko S., Itkonen O., Hämäläinen E.: Determination of testosterone in serum by liquid chromatography-tandem mass spectrometry . Scan J Clin Lab Inv 2008, 68, 50-57.
  • 21. Wang S., Lin S., Du L., Zhuang H., Flow injection chemiluminescence immunoassay for 17β-estradiol using an immunoaffinity column. Anal Bioanal Chem 2006, 384, 1186-1190.
  • 22. Witorsch R.J.: Endocrine dispruptors: Can biological effects and environmental risk be predicted? Regul Toxicol Pharmacol 2002, 36, 118-130.
  • 23. Woźniak В., Wojtoń В.: Immunoenzymatic method for progesterone detection in bovine urine, muscles, and serum by means of a comercial EIA test used in human medicine. Bull Vet Inst Pulawy 1995, 39, 139-145.
  • 24. Wu H., Ramsay C., Ozaeta P., Liu L., Aboleneen H.: Serum estradiol quantified by isotope dilution-gas chromatography/mass spectrometry. Clin Chem 2002, 48, 364-366.
  • 25. Yamada H., Yoshizawa K., Hayase T.: Sensitive determination method of estradiol in plasma using high-performance liquid chromatography with electrochemical detection. J Chromatogr В 2002, 775, 209-2013.

Uwagi

Rekord w opracowaniu

Typ dokumentu

Bibliografia

Identyfikatory

Identyfikator YADDA

bwmeta1.element.agro-5a42298b-48e4-4b20-b126-d13fdab0adf0
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.